How to recalibrate the RS350?

The calibrators RS350 are leaving our factory with a traceable calibration certificate.
All metrological instruments need to be recalibrated regular. The recalibration period depends on country regulations. In most countries portable calibrators have a recalibration period of one year.
The recalibration can be done in any for this purpose accredited laboratory.
The general principle is to compare the calibrator pulses with the reference standard pulses. Connect the P-OUT to the pulse input and GND to the ground, calculate the meter constant for the actual load and verify the RS350 accuracy with the higher accurate standard.

RS350 meter constants
RS350 meter constants

How to recalibrate the RS350?